Emerson - Rosemount 248HANAN0NSQ4 Head-Mount Temperature Transmitter – HART 4–20 mA, No Approval, No Enclosure, No Sensor, 3-Point Calibration Certificate

Emerson


Rosemount 248HANAN0NSQ4 Head-Mount Temperature Transmitter – HART 4–20 mA, No Approval, No Enclosure, No Sensor, 3-Point Calibration Certificate - 248HANAN0NSQ4

Product Rosemount 248HANAN0NSQ4 Head-Mount Temperature Transmitter – HART 4–20 mA, No Approval, No Enclosure, No Sensor, 3-Point Calibration Certificate
Brand Emerson
Product Price 0 LKR
Datasheet Download

Description

General
The Rosemount 248HANAN0NSQ4 is a DIN B head-mount temperature transmitter manufactured by Emerson. It is designed for reliable single-point temperature measurement in industrial process applications, converting compatible temperature sensor signals into a two-wire 4–20 mA output with HART digital communication.

This configuration is supplied without product approval, enclosure or temperature sensor. It includes a Q4 three-point calibration certificate, providing documented calibration at 0%, 50% and 100% of the configured range. The NEX Instrument listing specifies a calibration range of 0–150 °C.

Features
- DIN B head-mount temperature transmitter design.
- Two-wire 4–20 mA output with HART digital communication.
- Universal input for RTD, thermocouple, resistance, potentiometer and millivolt signals.
- Single-sensor measurement capability.
- Supplied without an enclosure.
- Supplied without a temperature sensor.
- No product approval included.
- Three-point calibration certificate (Q4).
- Open- and short-sensor diagnostics.
- Automatic self-calibration.
- Transmitter-sensor matching using Callendar–Van Dusen constants.
- SIL 2 capability according to IEC 61508 certification.

Key Specifications
- Brand: Rosemount
- Manufacturer: Emerson
- Model: 248
- Manufacturer Part Number: 248HANAN0NSQ4
- Product Type: Head-Mount Temperature Transmitter
- Transmitter Type: DIN B Head Mount
- Output Signal: Two-wire 4–20 mA with HART digital communication
- Sensor Input: RTD, thermocouple, resistance, potentiometer and bipolar millivolt
- Sensor Configuration: Single sensor
- Product Certification: NA – No approval
- Enclosure: N – No enclosure
- Conduit Entry: 0 – No enclosure
- Sensor Assembly Option: NS – No sensor
- Calibration Certificate: Q4 – Three-point calibration certificate
- Listed Calibration Range: 0–150 °C (supplier-listed configuration)
- Calibration Points: 0%, 50% and 100% of configured range
- Power Supply: 12.0–42.4 V DC at transmitter terminals
- Minimum Supply Voltage: 17.75 V DC with 250 Ω load
- Operating Ambient Temperature: −40 to +85 °C
- Storage Temperature: −50 to +120 °C
- Humidity: 0–95% RH, non-condensing
- Update Rate: Less than 0.5 seconds
- Damping: 5 seconds default; adjustable up to 32 seconds
- Turn-On Time: Less than 5 seconds with zero damping
- Input/Output Isolation: Tested to 500 V AC RMS
- Transmitter Weight: Approximately 50 g

Contacts and Delivery

FieldInstruments.lk
50/3, Angampitiya Rd, Ethul Kotte, Sri Lanka. 10100.
Phone: +94718555888
WhatsApp: +94 71 388 7888
Email: info@fieldinstruments.lk